Calibrating the Built-in Microphone
Calibrating the mic
- If you already know the sensitivity of your built-in microphone (in Pa/FS), you can
enter it directly in the Input Sensitivity text box of the calibration view and
skip the rest of the calibration procedure.
- Apply a known sound level to the built-in microphone of your iPhone.
- You will need to calibrate your iPhone's built-in microphone relative to a calibrated sound
level meter. You can produce the reference signal with SignalScope Pro's built-in tone generator
output connected to a loudspeaker (with a suitable amplifier in between, of course).
You can then measure the sound level at a fixed location with your calibrated meter,
remove the meter, and place
your iPhone's microphone as close to the same measurement location
as possible (and with the same orientation).
- A 1 kHz tone at 94 dB is typically used for microphone calibration.
- Performing a relative microphone calibration in a quiet, non-reverberant location is
recommended.
- Make sure the tone generator is turned on (in the Sig
Gen tab) before proceeding to the next step.
It might be a good idea to check the input signal in the
Oscilloscope display to be sure the signal looks good and that there is no visible distortion of
the waveform.
- Open the options view by tapping its icon in the toolbar.
- Tap I/O Device Options.
- In the top row of the Input Options table, make sure the desired input device is selected
(its name will be shown in parentheses).
- Make sure the Device Units selection is set to Pa (pascals).
- If you're using SignalScope Pro's built-in tone generator for a reference source, make sure
Audio Play Through is switched off.
- Tap Calibrate.
- The Calibration view displays the current measured input level as a text value as well as
in a horizontal bar meter. If the current measured input level agrees with the reference sound level
you have applied to the microphone (and there's no indication of input clipping), then you're done--your
microphone already has the correct sensitivity applied.
- Make sure your input is not clipping (you should not see the word "clip" displayed with
a red box around it in the upper left-hand corner of the screen).
If it is, you will need to decrease the level of your reference input signal. When the input
signal is no longer clipping, the red box around the word "clip" will disappear.
- Enter the reference sound level into the Ref. Input Level text box.
- Click the Calibrate button.
- Confirm that you would like to Calibrate.
- Now check to see that the measured input level closely matches the level of the
reference signal. If it does, your microphone is properly calibrated.
- Clicking the Calibrate button causes SignalScope Pro to automatically calculate a proper
sensitivity (of pascals relative to full scale input, or Pa/FS) for the microphone, based
on the current measured input level and the Ref. Input Level value.
- When working with the built-in microphone, the channel sensitivity and the device sensitivity
are always the same.